Main navigation

Sciemetric
  • Benchtop configuration 9” W x 11.25” H x 15” D
  • Powerful 32-bit processor & 24 –bit A/D converter
  • Language neutral color LCD 12 line 480 x 272 operator display
  • XY Graphing (pressure, loss, or flow)-English, Chinese, Korean, Spanish
  • Light ring indicators per test port identifies test port status Pass/Fail/In-test
  • Modular pneumatic and control packages
  • Expanded instrument control through EtherNet/IP™ Communication
  • Single test channel, with multiple test port options(maximum of 4)
  • Barcode ready input RS232 Alpha-numeric string stored with test result.
  • Auto setup, verifies test pressure and selects the best test timers for optimum results in the allowed time cycle (US Patent #7,594,424 B2)
  • Auto Calibration includes mass flow leak standard certified for pressure and flow
  • Transducer zero and span are automatically reset during auto calibration
  • Optional “External leak Standard” port 
  • “Performance factor” calibration feedback ensures accurate results
  • Manual edit of Calibration Program permitted
  • Self Test isolates the test instrument pneumatics to verify a “no leak” condition
  • Features ensure accurate results and easy operator setup and troubleshooting
  • Firmware validation document included

 Standard pressure transducer, valve manifolds, and regulator ranges:
 -14.5, -2, 0.5, 2, 10, 30, 100, 200, 400, 500 or electronic regulation (PSIG)

  • Optional low volume test manifold available
  • Absolute pressure transducer with accuracy of +/-0.05% of full scale
  • Six point pressure transducer calibration
  • Data resolution 0.00001 psi/ and scc/m.
  • Optional electronic regulator available with multi-point calibration
  • Optional internal vacuum generator available
  • Optional automatic supply shutoff available

CTS will optimize the flow sensor & flow range per application. Contact Application Engineering for available mass flow sensors.

PSIG, PSIV, inH2O, cmH2O, mmH2O, kg/cm2, ATM, inHg, cmHg, mmHg, kPa, Pa, MPa, Bar, mBar, Torr

Sccm, sccs, scch, slpm, slps, slph, scfm, scfs, scfh

Sec, msec, min

Units of measure can be instrument global or per test program
 

  • Expanded control through EtherNet/IP™ Communication
  • 6 inputs/6 outputs with up to five programmable tooling motions with feedback
  • Optional “CTS Connect” pneumatic actuator port (5 pneumatic outputs maximum)
  • Tooling I/O is  externally connected
  • I/O 24 VDC – optional I/O cable “Cable,CTS, P56-540”
  • Up to 5000 LIFO part results including Date / Time / Barcode / Test Results
  • Selectable part serialization in the test data stream
  • SPC data tracking for part traceability
  • Expandable download through USB port
  • Key lock or password protected program and data security
  • 88-264 VAC, 47-63 Hz or optional 22-26 VDC
  • RS232(1)
  • Telnet(1)
  • Email Alerts
  • Barcode ready (RS232)
  • Mobile text
  • USB mass storage device for backup, restore, and reports (FAT32, USB1.0,1.1,2.0)
  • 1/8” NPT, Male Luer or Female Luer fitting for test line attachment

Multi-test Configurable for Pressure Decay, Vacuum Decay or Mass Flow Test Requirements  

  • Pressure/Vacuum decay-pressure change
  • Pressure/Vacuum decay-with leak standard
  • Pressure/Vacuum decay-pressure change/time
  • Pressure decay-differential pressure loss
  • Occlusion test (vacuum or pressure)
  • Mass flow
  • Mass flow continuous direct flow reading
  • Mass flow with certified mass flow leak standard
  • Ramp up or down to change in pressure event
  • Ramp to flow event with direct flow measurement
  • Ramp to “Proof” test
  • Sequential Test Linking
  • “Parent Program” holds a customized test sequence for your multi-test application
  • Storage for 99 pre-programmed test sequences
  • Test Program can be selected through keypad, RS232, or Binary
  • Test results evaluation is Selectable (Fail/Pass/Fail etc)
  • Test result output format is configurable format

Selectable Test Parameters

  •  Quick Test
  •  Test Pressure Level Compensation (selectable on/off)
  • Adjustable Environmental Drift Correction per test program

REQUEST A Quote