Advanced Product Features
- Wall mount configuration 9” W x 6.5” H x 7.25” D
- Powerful 32-bit processor & 24 –bit A/D converter
- Language neutral color LCD 6 line 480 x 136 operator display
- Bar Graph display (pressure, loss, or flow)-English, Chinese, Korean, Spanish
- Individual LED Pass/Fail/In-test test status indicators
- Modular pneumatic and control packages
- Single test channel
Standard Features
- Auto Calibration includes mass flow leak standard certified for pressure and flow
- Transducer zero and span are automatically reset during auto calibration
- “Performance factor” calibration feedback ensures accurate results
- Manual edit of Calibration Program permitted
- Self Test isolates the test instrument pneumatics to verify a “no leak” condition
- Features ensure accurate results and easy operator setup and troubleshooting
- Firmware validation document included
Standard Test Pressure Ranges
Standard pressure transducer, valve manifolds, and regulator ranges:
-14.5, 15, 100 (PSIG)
- Optional High cV flow valves and manifold available up to 100 psig range allows 5 times faster fill and exhaust to reduce cycle times on large parts
- Absolute pressure transducer
- Two (2) point pressure calibration
- Data resolution 0.00001 psi/ and scc/m.
Selectable Test Units of Measure
- Pressure: PSIG, PSIV, inH2O, cmH2O, mmH2O, kg/cm2, ATM, inHg, cmHg, mmHg, kPa, Pa, MPa, Bar, mBar, Torr
- Flow: Sccm, sccs, scch, slpm, slps, slph, scfm, scfs, scfh
- Time: Sec, msec, min
- Units of measure can be instrument global or per test program
Tooling Control
- 6 inputs/3 outputs- 1in/1out “seal only” tooling motion - no feedback
- Tooling I/O is internally connected
- I/O 24 VDC – optional I/O cable “Cable,CTS, P56-540”
Test Data Storage
- Up to 5000 LIFO part results including Date / Time / Test Results
- SPC data tracking for part traceability
Electrical Requirements and Connectivity
- 88-264 VAC, 47-63 Hz or optional 22-26 VDC
- RS232(1)
- ¼” Colder fitting for test line attachment
Testing Technologies
Single Test Configurable for Pressure or Vacuum Decay Test Requirements
- Pressure/Vacuum decay-pressure change
- Pressure/Vacuum decay-with certified mass flow leak standard
- Pressure/Vacuum decay-pressure change/time
- Storage for 4 pre-programmed test sequences
- Test Program can be selected through keypad or binary input
- Test results evaluation is fixed (Fail/Pass/Fail)
- Test result output format is fixed
Selectable Test Parameters
- Quick Test
- Test Pressure Level Compensation
- Adjustable Environmental Drift Correction per test program