Main navigation

Sciemetric

The Sentinel C20WE (World Edition) is a high performance, low cost, quality pressure decay leak detection instrument. Its small packaging enables the C20WE to integrate easily into product line testing applications. The Sentinel C20WE is easy to configure. The high flow valve option design reduces cycle times by filling and exhausting the part up to five times faster. It is excellent for high speed and/or big volume applications.

"Quik Test" feature, incorporated into the Sentinel C20WE, provides a highly reliable test on all parts while shortening the test portion of the cycle time by up to 80%. Throughput on existing test systems can be increased without major investment in fixtures, tooling or operators. On systems which are already running shortened cycle times for all parts, the risk of accepting marginal reject parts is reduced.

  • Wall mount configuration 9” W x 6.5” H x 7.25” D
  • Powerful 32-bit processor & 24 –bit A/D converter
  • Language neutral color LCD 6 line 480 x 136 operator display
  • Bar Graph display  (pressure, loss, or flow)-English, Chinese, Korean, Spanish
  • Individual LED Pass/Fail/In-test test status indicators
  • Modular pneumatic and control packages
  • Single test channel
  • Auto Calibration includes mass flow leak standard certified for pressure and flow
  • Transducer zero and span are automatically reset during auto calibration
  • “Performance factor” calibration feedback ensures accurate results
  • Manual edit of Calibration Program permitted
  • Self Test isolates the test instrument pneumatics to verify a “no leak” condition
  • Features ensure accurate results and easy operator setup and troubleshooting
  • Firmware validation document included


Standard pressure transducer, valve manifolds, and regulator ranges:

-14.5, 15, 100 (PSIG)

  • Optional High cV flow valves and manifold available up to 100 psig range allows 5 times faster fill and exhaust to reduce cycle times on large parts
  • Absolute pressure transducer
  • Two (2) point pressure calibration
  • Data resolution 0.00001 psi/ and scc/m.
  • Pressure: PSIG, PSIV, inH2O, cmH2O, mmH2O, kg/cm2, ATM, inHg, cmHg, mmHg, kPa,  Pa, MPa, Bar,  mBar, Torr
  • Flow: Sccm, sccs,  scch,  slpm,  slps,  slph, scfm, scfs, scfh
  • Time: Sec, msec, min
  • Units of measure can be instrument global or per test program
  • 6 inputs/3 outputs- 1in/1out  “seal only” tooling motion - no feedback
  • Tooling I/O is  internally connected
  • I/O 24 VDC – optional I/O cable “Cable,CTS, P56-540”
  • Up to 5000 LIFO part results including Date / Time / Test Results
  • SPC data tracking for part traceability
  • 88-264 VAC, 47-63 Hz or optional 22-26 VDC
  • RS232(1)
  • ¼” Colder fitting for test line attachment

Single Test Configurable for Pressure or Vacuum Decay Test Requirements

  • Pressure/Vacuum decay-pressure change
  • Pressure/Vacuum decay-with certified mass flow leak standard
  • Pressure/Vacuum decay-pressure change/time
  • Storage for 4 pre-programmed test sequences
  • Test Program can be selected through keypad or binary input
  • Test results evaluation is fixed (Fail/Pass/Fail)
  • Test result output format is fixed

Selectable Test Parameters

  • Quick Test
  • Test Pressure Level Compensation
  • Adjustable Environmental Drift Correction per test program

REQUEST A Quote